Comparison of two methods for measurement of equine insulin

Diagnosis of equine hyperinsulinemia requires an accurate method for quantification of equine insulin concentrations. The objectives of the current study were to compare 2 commercially available techniques for measurement of equine insulin, the radioimmunoassay (RIA) and chemiluminescent immunoassay...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of veterinary diagnostic investigation 2014-07, Vol.26 (4), p.527-530
Hauptverfasser: Banse, Heidi E., McCann, Joseph, Yang, Fan, Wagg, Catherine, McFarlane, Dianne
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Diagnosis of equine hyperinsulinemia requires an accurate method for quantification of equine insulin concentrations. The objectives of the current study were to compare 2 commercially available techniques for measurement of equine insulin, the radioimmunoassay (RIA) and chemiluminescent immunoassay (CIA). Recovery was poor for both assays, but worse for the CIA. Serial dilution of a high endogenous insulin sample yielded better linearity for the RIA (r2 = 0.99, P < 0.001) than the CIA (r2 = 0.92, P = 0.009). Bland–Altman analysis indicated that the CIA was, on average, 91 pmol/l higher than the RIA, with wide limits of agreement (95% limits of agreement: –508 to 691 pmol/l). These findings suggest that results between the assays should not be considered interchangeable.
ISSN:1040-6387
1943-4936