Temperature effects on the performance of RuO sub(2) thin-film pH sensor

A miniaturized pH sensor structure employing an R.F. sputtered ruthenium oxide sensing thin-film electrode in conjunction with an electroplated Ag/AgCl reference electrode is developed and the effect of the temperature of the pH test solution on the performance of the sensor is investigated. Experim...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors and actuators. A. Physical. 2015-09, Vol.233, p.414-421
Hauptverfasser: Sardarinejad, A, Maurya, D K, Khaled, M, Alameh, K
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A miniaturized pH sensor structure employing an R.F. sputtered ruthenium oxide sensing thin-film electrode in conjunction with an electroplated Ag/AgCl reference electrode is developed and the effect of the temperature of the pH test solution on the performance of the sensor is investigated. Experimental results show that the pH sensor employing 300 nm thick RuO sub(2) sensing electrode exhibits Nernstian slopes of 65.66 mV/pH, 69.16 mV/pH, 72.50 mV/pH, 78.50 mV/pH and 84.50 mV/pH for pH test solution temperatures of 1.5 degree C, 12 degree C, 22 degree C, 37 degree C and 50 degree C, respectively. The performance of the pH sensor in terms of sensitivity, response time, stability and reversibility is experimentally evaluated.
ISSN:0924-4247
DOI:10.1016/j.sna.2015.07.027