Does high hip center decrease range of motion in total hip arthroplasty? A computer simulation study

Abstract Background High placement of a cementless acetabular cup is often required to have sufficient bone coverage in hip dysplasia. We examined whether a high hip center decreases the postoperative range of motion (ROM) in total hip arthroplasty (THA). Methods Using a computer software, ROM and b...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2016-10, Vol.31 (10), p.2342-2347
Hauptverfasser: Komiyama, Keisuke, MD, Nakashima, Yasuharu, MD, PhD, Hirata, Masanobu, MD, PhD, Hara, Daisuke, MD, Kohno, Yusuke, MD, Iwamoto, Yukihide, MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Background High placement of a cementless acetabular cup is often required to have sufficient bone coverage in hip dysplasia. We examined whether a high hip center decreases the postoperative range of motion (ROM) in total hip arthroplasty (THA). Methods Using a computer software, ROM and bone coverage were examined in 32 patients with unilateral osteoarthritis of the hip with Crowe type II or III hip dysplasia. The cup was placed at the anatomic hip center and moved vertically in 5-mm increments. Vertical center of rotation (V-COR) was defined as the distance from the head center to the interteardrop line. The required ROM was defined as flexion ≥ 110°, internal rotation at 90° flexion (IR) ≥ 30°, extension ≥ 30°, and external rotation (ER) ≥ 30°. We determined the V-COR to satisfy the required ROM and Cup center-edge angle (Cup-CE) ≥ 0°. Results Only 40.6% of the patients satisfied Cup-CE ≥ 0° at the anatomic hip center. Bone coverage increased with the peak at 30, 35 mm of V-COR (90.6% satisfied Cup-CE ≥ 0°) and decreased at ≥ 40 mm. Higher V-COR resulted in lower range of flexion and IR, but increased the range of extension and ER. The cut-off value from the receiver operating characteristic (ROC) curve for both flexion and IR was a V-COR of 35 mm. Conclusion The higher hip center gained more bone coverage but decreased the range of hip flexion and IR. Acceptable V-COR in hip dysplasia was around 35 mm, about 20 mm above the true hip center, in terms of ROM.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2016.03.014