Techniques for pollution monitoring in remote sites. II: Ozone
An O sub(3) monitoring system for use in remote sites lacking AC line power was developed. A chemiluminescent O sub(3) monitor, ethylene delivery system, and a DC power supply were combined to allow unattended O sub(3) measurements over extended periods of time. A Campbell Scientific 21X micrologger...
Gespeichert in:
Veröffentlicht in: | Water, air, and soil pollution air, and soil pollution, 1993-01, Vol.68 (3-4), p.391-397 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An O sub(3) monitoring system for use in remote sites lacking AC line power was developed. A chemiluminescent O sub(3) monitor, ethylene delivery system, and a DC power supply were combined to allow unattended O sub(3) measurements over extended periods of time. A Campbell Scientific 21X micrologger in combination with a relay was used to control the ethylene delivery solenoid and O sub(3) analyzer power, and to record results. The system was tested over a six week period in late summer 1990 and gave results comparable to those from surrounding sites. |
---|---|
ISSN: | 0049-6979 1573-2932 |
DOI: | 10.1007/BF00478465 |