Angular vibrations of cryogenically cooled double-crystal monochromators

The effect of angular vibrations of the crystals in cryogenically cooled monochromators on the beam performance has been studied theoretically and experimentally. A simple relation between amplitude of the vibrations and size of the focused beam is developed. It is shown that the double‐crystal mono...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of synchrotron radiation 2016-09, Vol.23 (5), p.1097-1103
Hauptverfasser: Sergueev, I., Döhrmann, R., Horbach, J., Heuer, J.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The effect of angular vibrations of the crystals in cryogenically cooled monochromators on the beam performance has been studied theoretically and experimentally. A simple relation between amplitude of the vibrations and size of the focused beam is developed. It is shown that the double‐crystal monochromator vibrations affect not only the image size but also the image position along the optical axis. Several methods to measure vibrations with the X‐ray beam are explained and analyzed. The methods have been applied to systematically study angular crystal vibrations at monochromators installed at the PETRA III light source. Characteristic values of the amplitudes of angular vibrations for different monochromators are presented. A theoretical and experimental investigation of the effect of angular vibrations of a double‐crystal monochromator on the X‐ray beam quality is presented.
ISSN:1600-5775
1600-5775
DOI:10.1107/S1600577516011188