Transcatheter Valve-in-Valve: A Cautionary Tale

Transcatheter aortic valve replacement (TAVR) by valve-in-valve (VIV) implantation is an alternative treatment for high-risk patients with a degenerating aortic bioprosthesis. We present a case of transapical TAVR VIV with a 29-mm Edwards SAPIEN XT (ESV) (Edwards Lifesciences, Irvine, CA) into a 29-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Annals of thoracic surgery 2016-09, Vol.102 (3), p.e211-e213
Hauptverfasser: Luc, Jessica G.Y, Shanks, Miriam, MD, PhD, Tyrrell, Benjamin D., MD, Welsh, Robert C., MD, Butler, Craig R., MD, MS, Meyer, Steven R., MD, PhD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Transcatheter aortic valve replacement (TAVR) by valve-in-valve (VIV) implantation is an alternative treatment for high-risk patients with a degenerating aortic bioprosthesis. We present a case of transapical TAVR VIV with a 29-mm Edwards SAPIEN XT (ESV) (Edwards Lifesciences, Irvine, CA) into a 29-mm Medtronic Freestyle stentless bioprosthesis (Medtronic Inc, Minneapolis, MN) in which unanticipated dilatation of the Freestyle bioprosthesis resulted in intraprocedural embolization of the TAVR valve, necessitating urgent conversion to a conventional surgical aortic valve replacement (AVR). Our experience suggests that TAVR VIV with the 29-mm ESV in the setting of a degenerated 29-mm Freestyle stentless bioprosthesis must be undertaken with caution.