Evaluation of 2 portable ion-selective electrode meters for determining whole blood, plasma, urine, milk, and abomasal fluid potassium concentrations in dairy cattle

Two low-cost ion-selective electrode (ISE) handheld meters (CARDY C-131, LAQUAtwin B-731; Horiba Ltd., Albany, NY) have recently become available for measuring the potassium concentration ([K+]) in biological fluids. The primary objective of this study was to characterize the analytical performance...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of dairy science 2016-09, Vol.99 (9), p.7330-7343
Hauptverfasser: Megahed, A.A., Hiew, M.W.H., Grünberg, W., Constable, P.D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two low-cost ion-selective electrode (ISE) handheld meters (CARDY C-131, LAQUAtwin B-731; Horiba Ltd., Albany, NY) have recently become available for measuring the potassium concentration ([K+]) in biological fluids. The primary objective of this study was to characterize the analytical performance of the ISE meters in measuring [K+] in bovine whole blood, plasma, urine, milk, and abomasal fluid. We completed 6 method comparison studies using 369 whole blood and plasma samples from 106 healthy periparturient Holstein-Friesian cows, 138 plasma samples from 27 periparturient Holstein-Friesian cows, 92 milk samples and 204 urine samples from 16 lactating Holstein-Friesian cows, and 94 abomasal fluid samples from 6 male Holstein-Friesian calves. Deming regression and Bland-Altman plots were used to characterize meter performance against reference methods (indirect ISE, Hitachi 911 and 917; inductively coupled plasma-optical emission spectroscopy). The CARDY ISE meter applied directly in plasma measured [K+] as being 7.3% lower than the indirect ISE reference method, consistent with the recommended adjustment of +7.5% when indirect ISE methods are used to analyze plasma. The LAQUAtwin ISE meter run in direct mode measured fat-free milk [K+] as being 3.6% lower than the indirect ISE reference method, consistent with a herd milk protein percentage of 3.4%. The LAQUAtwin ISE meter accurately measured abomasal fluid [K+] compared to the indirect ISE reference method. The LAQUAtwin ISE meter accurately measured urine [K+] compared to the indirect ISE reference method, but the median measured value for urine [K+] was 83% of the true value measured by inductively coupled plasma-optical emission spectroscopy. We conclude that the CARDY and LAQUAtwin ISE meters are practical, low-cost, rapid, accurate point-of-care instruments suitable for measuring [K+] in whole blood, plasma, milk, and abomasal fluid samples from cattle. Ion-selective electrode methodology is not suitable for measuring [K+] in bovine urine.
ISSN:0022-0302
1525-3198
DOI:10.3168/jds.2015-10821