Small field correction factors for the IBA Razor

Highlights • The IBA Razor overresponds in small fields as with other silicon diodes. • Correction factors for the overresponse are presented for Varian linacs. • The correction factor of FF beams was compared to that of FFF beams. • The overresponse can be predicted mathematically based on the fiel...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physica medica 2016-08, Vol.32 (8), p.1025-1029
Hauptverfasser: Liu, Paul Z.Y, Reggiori, Giacomo, Lobefalo, Francesca, Mancosu, Pietro, Tomatis, Stefano, McKenzie, David R, Suchowerska, Natalka
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • The IBA Razor overresponds in small fields as with other silicon diodes. • Correction factors for the overresponse are presented for Varian linacs. • The correction factor of FF beams was compared to that of FFF beams. • The overresponse can be predicted mathematically based on the field size.
ISSN:1120-1797
1724-191X