Small field correction factors for the IBA Razor
Highlights • The IBA Razor overresponds in small fields as with other silicon diodes. • Correction factors for the overresponse are presented for Varian linacs. • The correction factor of FF beams was compared to that of FFF beams. • The overresponse can be predicted mathematically based on the fiel...
Gespeichert in:
Veröffentlicht in: | Physica medica 2016-08, Vol.32 (8), p.1025-1029 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Highlights • The IBA Razor overresponds in small fields as with other silicon diodes. • Correction factors for the overresponse are presented for Varian linacs. • The correction factor of FF beams was compared to that of FFF beams. • The overresponse can be predicted mathematically based on the field size. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2016.07.004 |