A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system for the measurement of total gaseous sulfur species in air
A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system with a flameless, temperature-controlled furnace combustion source was developed for real-time measurement of total sulfur gases in air. This measurement system demonstrated a linear dynamic range exceeding five orders of magnitu...
Gespeichert in:
Veröffentlicht in: | Atmospheric environment (1994) 1999-02, Vol.33 (4), p.625-632 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A continuous monitor-sulfur chemiluminescence detector (CM-SCD) system with a flameless, temperature-controlled furnace combustion source was developed for real-time measurement of total sulfur gases in air. This measurement system demonstrated a linear dynamic range exceeding five orders of magnitude and equimolar sensitivity to the most prevalent atmospheric sulfur gases. A detection limit of 10
pptv was obtained using 10
min signal averages. On a real-time basis, detection in the 20–50
pptv range was demonstrated. After modification of its sample inlet system, the CM-SCD showed no appreciable interference effects from the addition of H
2O vapor, NO
2 or O
3 to the sample matrix. In the recent Gas-Phase Sulfur Intercomparison Experiment (GASIE), the CM-SCD compared favorably with SO
2 measurement methods. |
---|---|
ISSN: | 1352-2310 1873-2844 |
DOI: | 10.1016/S1352-2310(98)00122-8 |