Boston Type 1 Keratoprosthesis: Visual Outcomes, Device Retention, and Complications
PURPOSE:To determine the visual outcomes, device retention, and complications after Boston type 1 keratoprosthesis (KPro-1) device implantation. METHODS:Comprehensive review of every case of KPro-1 implantation at a tertiary eye care center. RESULTS:The initial KPro-1 procedure, performed in 75 eyes...
Gespeichert in:
Veröffentlicht in: | Cornea 2016-09, Vol.35 (9), p.1165-1174 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | PURPOSE:To determine the visual outcomes, device retention, and complications after Boston type 1 keratoprosthesis (KPro-1) device implantation.
METHODS:Comprehensive review of every case of KPro-1 implantation at a tertiary eye care center.
RESULTS:The initial KPro-1 procedure, performed in 75 eyes without a previous keratoprosthesis, was included in the analysis. During the first 6 postoperative months, improvement occurred in the mean preoperative best-corrected visual acuity of 20/1265 to a mean best-obtained postoperative visual acuity of 20/97 (P < 0.001). After a mean follow-up period of 41.4 months (range, 0.8–82.8 months), the final mean best-corrected visual acuity was 20/428. Improved vision was recorded in 43 eyes (57.3%), ambulatory vision (≥20/400) in 47 eyes (62.7%), intermediate functional vision (≥20/80) in 23 eyes (30.7%), and full functional vision (≥20/40) in 11 eyes (14.7%). The initial device was retained in 64 eyes (85.3%), with a Kaplan–Meier retention probability of 96% at 6 months and 82% at 5 years. One or more sight-threatening complications occurred in 51 eyes (68%). These included device extrusion in 11 eyes (14.7%), ulcerative keratitis in 12 eyes (16%), endophthalmitis in 7 eyes (9.3%), sterile vitritis in 3 eyes (4%), retroprosthetic membranes in 25 eyes (33.3%), maculopathy in 26 eyes (34.7%), retinal detachment in 9 eyes (12%), and progressive optic neuropathy in 7 eyes (9.3%).
CONCLUSIONS:Boston KPro-1 implantation is associated with satisfactory visual outcomes and excellent device retention in a majority of cases. However, serious postoperative complications are common and may compromise the final visual result. |
---|---|