Ex vivo simulation of cardiopulmonary bypass with human blood for hemocompatibility testing
Object: Experimental circuits for biomaterial surface testing are frequently limited by the tested blood volume, composition of the circuit, flow conditions and the use of animal blood. This report describes an ex vivo set-up for simulated cardiopulmonary bypass with human blood perfusion. We invest...
Gespeichert in:
Veröffentlicht in: | Perfusion 2016-07, Vol.31 (5), p.376-383 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Object:
Experimental circuits for biomaterial surface testing are frequently limited by the tested blood volume, composition of the circuit, flow conditions and the use of animal blood. This report describes an ex vivo set-up for simulated cardiopulmonary bypass with human blood perfusion. We investigated the clinical generalizability of the observed effects on hematological and metabolic parameters and the hemocompatibility of the system.
Methods:
The simulated cardiopulmonary bypass circuit consisted of a heparin-coated tubing system connected to an oxygenator and a venous reservoir. Normothermic flow of blood obtained from healthy donors was maintained at 2.4 L/min/m2 by a roller pump. Heparin was dosed to obtain a target activated clotting time (ACT) ⩾500 s. Blood was drawn at baseline and 0, 10, 60 and 120 minutes following the initiation of blood flow to determine hematological and metabolic parameters and the hemocompatibility of the extracorporeal system. Data were analyzed using repeated measures ANOVA.
Results:
Two hours of blood perfusion resulted in a small, but clinically unimportant reduction in hematocrit, whereas hemoglobin levels and red blood cell, platelet and leukocyte counts remained stable. There was a significant increase in ACT throughout the experiment. While pO2 levels and the pH remained unaltered during the experiment, pCO2 values decreased from 51 ± 6 mmHg at T0 to 41 ± 3 mmHg at T120 (p |
---|---|