Effectiveness of the implementation of a simple radiation reduction protocol in the catheterization laboratory
Abstract Background and purpose A reduction in radiation doses at the catheterization laboratory, maintaining the quality of procedures is essential. Our objective was to analyze the results of a simple radiation reduction protocol at a high-volume interventional cardiology unit. Methods We analyzed...
Gespeichert in:
Veröffentlicht in: | Cardiovascular revascularization medicine 2016-07, Vol.17 (5), p.328-332 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Background and purpose A reduction in radiation doses at the catheterization laboratory, maintaining the quality of procedures is essential. Our objective was to analyze the results of a simple radiation reduction protocol at a high-volume interventional cardiology unit. Methods We analyzed 1160 consecutive procedures: 580 performed before the implementation of the protocol and 580 after it. The protocol consisted in: the reduction of the number of ventriculographies and aortographies, the optimization of the colimation and the geometry of the X ray tube-patient-receptor, the use of low dose-rate fluoroscopy and the reduction of the number of cine sequences using the software “last fluoroscopy hold”. Results There were no significant differences in clinical baseline features nor in the procedural characteristics with the exception of a higher percentage of radial approach (30,7% vs 69.6%;p < 0.001) and of percutaneous coronary interventions of chronic total occlusions after the implementation of the protocol (2.1% vs 6.7%;p = 0,001). Angiographic success was similar during both periods (98.3% vs 99.2%;p = 0.2). There were no significant differences between both periods regarding the overall duration of the procedures (26.9 vs 29.6 min; p = 0.14), nor the fluoroscopy time (13.3 vs 13.2 min; p = 0.8). We observed a reduction in the percentage of procedures with ventriculography (80.9% vs 7.1%;p < 0.0001) or aortography (15.4% vs 4.4%;p < 0.0001), the cine runs (21.8 vs 6.9;p < 0.0001) and the dose-area product (165 vs 71 Gyxcm2 ;p < 0.0001). Conclusions With the implementation of a simple radiation reduction protocol, a 57% reduction of dose-area product was observed without a reduction in the quality or the complexity of procedures. |
---|---|
ISSN: | 1553-8389 1878-0938 |
DOI: | 10.1016/j.carrev.2016.03.012 |