Translational and rotational localization errors in cone-beam CT based image-guided lung stereotactic radiotherapy

Highlights • We determined localization and residual errors for 57 lung SBRT using CBCT. • Translational and rotational errors were detected using ITV/CBCT matching technique. • The automatic 6D correction with the robotic couch increased localization accuracy. • 5 mm safety margin to the ITV result...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physica medica 2016-07, Vol.32 (7), p.859-865
Hauptverfasser: Garibaldi, Cristina, Piperno, Gaia, Ferrari, Annamaria, Surgo, Alessia, Muto, Matteo, Ronchi, Sara, Bazani, Alessia, Pansini, Floriana, Cremonesi, Marta, Jereczek-Fossa, Barbara Alicja, Orecchia, Roberto
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • We determined localization and residual errors for 57 lung SBRT using CBCT. • Translational and rotational errors were detected using ITV/CBCT matching technique. • The automatic 6D correction with the robotic couch increased localization accuracy. • 5 mm safety margin to the ITV resulted adequate using the automatic 6D correction.
ISSN:1120-1797
1724-191X