Translational and rotational localization errors in cone-beam CT based image-guided lung stereotactic radiotherapy
Highlights • We determined localization and residual errors for 57 lung SBRT using CBCT. • Translational and rotational errors were detected using ITV/CBCT matching technique. • The automatic 6D correction with the robotic couch increased localization accuracy. • 5 mm safety margin to the ITV result...
Gespeichert in:
Veröffentlicht in: | Physica medica 2016-07, Vol.32 (7), p.859-865 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Highlights • We determined localization and residual errors for 57 lung SBRT using CBCT. • Translational and rotational errors were detected using ITV/CBCT matching technique. • The automatic 6D correction with the robotic couch increased localization accuracy. • 5 mm safety margin to the ITV resulted adequate using the automatic 6D correction. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2016.05.055 |