Non‐invasive continuous arterial pressure and pulse pressure variation measured with Nexfin® in patients following major upper abdominal surgery: a comparative study

Summary We compared the accuracy and precision of the non‐invasive Nexfin® device for determining systolic, diastolic, mean arterial pressure and pulse pressure variation, with arterial blood pressure values measured from a radial artery catheter in 19 patients following upper abdominal surgery. Mea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Anaesthesia 2016-07, Vol.71 (7), p.788-797
Hauptverfasser: Wilde, R. B. P., Wit, F., Geerts, B. F., Vliet, A. L., Aarts, L. P. H. J., Vuyk, J., Jansen, J. R. C.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Summary We compared the accuracy and precision of the non‐invasive Nexfin® device for determining systolic, diastolic, mean arterial pressure and pulse pressure variation, with arterial blood pressure values measured from a radial artery catheter in 19 patients following upper abdominal surgery. Measurements were taken at baseline and following fluid loading. Pooled data results of the arterial blood pressures showed no difference between the two measurement modalities. Bland–Altman analysis of pulse pressure variation showed significant differences between values obtained from the radial artery catheter and Nexfin finger cuff technology (mean (SD) 1.49 (2.09)%, p 
ISSN:0003-2409
1365-2044
DOI:10.1111/anae.13503