A patient immobilization device for prone breast radiotherapy: Dosimetric effects and inclusion in the treatment planning system

Highlights • Beams intersecting the AIO patient positioning device are attenuated considerably. • Build-up doses increase substantially in beams intersecting the positioning device. • Attenuation is accurately modeled by inclusion of the device in the planning CT. • This approach fails to precisely...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physica medica 2016-06, Vol.32 (6), p.758-766
Hauptverfasser: De Puysseleyr, A, De Neve, W, De Wagter, C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Beams intersecting the AIO patient positioning device are attenuated considerably. • Build-up doses increase substantially in beams intersecting the positioning device. • Attenuation is accurately modeled by inclusion of the device in the planning CT. • This approach fails to precisely predict build-up doses.
ISSN:1120-1797
1724-191X
DOI:10.1016/j.ejmp.2016.04.013