Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films
A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthr...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2015-07, Vol.165 (1-4), p.244-249 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations. |
---|---|