Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films

A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation protection dosimetry 2015-07, Vol.165 (1-4), p.244-249
Hauptverfasser: Bordier, C, Klausz, R, Desponds, L
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations.