Examination of the relevance of using radiochromic films in measuring entrance skin dose distribution in conventional digital mammography
Based on manufacturer specifications, radiochromic films are sensitive enough to be used for dosimetry in digital mammography (DM). The aim of this work was to study the feasibility of measuring entrance surface dose (ESD) distribution using Gafchromic XR-QA2 films. The films were irradiated followi...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2015-07, Vol.165 (1-4), p.373-375 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Based on manufacturer specifications, radiochromic films are sensitive enough to be used for dosimetry in digital mammography (DM). The aim of this work was to study the feasibility of measuring entrance surface dose (ESD) distribution using Gafchromic XR-QA2 films. The films were irradiated following a standard clinical two-view screening mammography protocol using a full-field digital mammography (FFDM) imaging system. The films were then digitised using a flatbed scanner. The calibration curve relating the readings from a calibrated ionisation chamber and the films' net optical density (NOD) could not be obtained. The examination of the calibration data revealed non-sensitivity of the films to resolve dose differences below 20 mGy at 28 kVp. Therefore, radiochromic films were found not to be suitable for measuring ESD profiles in DM. A 2D map of the NOD of the irradiated films obtained using in-house developed MATLAB computer program is presented. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncv126 |