An HPLC Method for the Determination of Nifekalant Hydrochloride in Canine Plasma and its Application to a Pharmacokinetic Study

In this study, a simple and selective high-performance liquid chromatography method was developed and validated for the determination of nifekalant hydrochloride in canine plasma. Liquid-liquid extraction was used to separate nifekalant hydrochloride from canine plasma and the mean extraction recove...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatographic science 2013-10, Vol.51 (9), p.867-871
Hauptverfasser: Zhou, Yi, Liu, Xiu-Mei, Wang, Ling, Jiang, Xue-Hua
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, a simple and selective high-performance liquid chromatography method was developed and validated for the determination of nifekalant hydrochloride in canine plasma. Liquid-liquid extraction was used to separate nifekalant hydrochloride from canine plasma and the mean extraction recoveries of nifekalant hydrochloride and the internal standard were 82.31 and 94.81%, respectively. The chromatographic separation was performed on a Dikma Diamonsil column with a mobile phase consisting of acetonitrile-20mM phosphate buffer (pH 6.2; 30:70, v/v) with flow rate of 1.0 mL/min. The standard curve was linear over the concentration range of 20-10,000 ng/mL (r 2 > 0.99). The intra-batch and inter-batch accuracy for nifekalant hydrochloride at four concentrations were 93.14-100.47% and 96.12-103.77%, respectively. The relative standard deviations were less than 15%. The method was successfully applied to a pharmacokinetic study after the intravenous administration of nifekalant hydrochloride to beagle dogs.
ISSN:0021-9665
1945-239X
DOI:10.1093/chromsci/bms182