Bicuspid Aortic Valve Stenosis: Favorable Early Outcomes With a Next-Generation Transcatheter Heart Valve in a Multicenter Study

This study evaluated the results of transcatheter aortic valve replacement (TAVR) in bicuspid aortic stenosis (AS) using a new-generation TAVR device. A bicuspid AS is often considered a relative contraindication to TAVR. Although initial reports have demonstrated feasibility using early-generation...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:JACC. Cardiovascular interventions 2016-04, Vol.9 (8), p.817-824
Hauptverfasser: Perlman, Gidon Y, Blanke, Philipp, Dvir, Danny, Pache, Gregor, Modine, Thomas, Barbanti, Marco, Holy, Erik W, Treede, Hendrik, Ruile, Philipp, Neumann, Franz-Josef, Gandolfo, Caterina, Saia, Francesco, Tamburino, Corrado, Mak, George, Thompson, Christopher, Wood, David, Leipsic, Jonathon, Webb, John G
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study evaluated the results of transcatheter aortic valve replacement (TAVR) in bicuspid aortic stenosis (AS) using a new-generation TAVR device. A bicuspid AS is often considered a relative contraindication to TAVR. Although initial reports have demonstrated feasibility using early-generation devices, outcomes have not matched those seen with tricuspid AS. Paravalvular aortic regurgitation (AR) has been particularly problematic. We collected baseline characteristics, procedural data, and 30-day clinical follow-up findings from 8 centers in Europe and Canada that had performed TAVR in bicuspid AS using the SAPIEN 3 valve. 51 patients underwent TAVR using the SAPIEN 3 valve. Patient mean age was 76.2 ± 9.3 years and the Society of Thoracic Surgeons predicted risk of mortality scores were 5.2 ± 3.7%. Bicuspid valve types were: type 0, 11.8%; type 1, 82.3%; and type 2, 1.9%. There were no cases of valve embolization or need for a second valve. Post-dilation was performed in 7.8%. The mean aortic gradient decreased from 49.4 ± 16.0 mm Hg to 11.2 ± 4.7 mm Hg. Post-implantation AR was none/trivial in 63% and mild in 37%. There were no cases of moderate or severe AR. At 30-day follow-up, there were 2 deaths (3.9%), 2 major vascular complications, and 12 patients (23.5%) required pacemaker implantation. TAVR in bicuspid AS using a new-generation device was feasible and effective with favorable valve performance and no cases of moderate or severe AR.
ISSN:1876-7605
DOI:10.1016/j.jcin.2016.01.002