Type IIIb Endoleak and Relining: A Mathematical Model of Distraction Forces
Purpose: To examine the changes in distraction force following relining of a conventional abdominal aortic stent-graft with a type IIIb endoleak using the Nellix endovascular sealing device compared to a unilateral stent-graft. Methods: Relining is often used to repair type IIIb endoleaks, but the c...
Gespeichert in:
Veröffentlicht in: | Journal of endovascular therapy 2016-04, Vol.23 (2), p.297-301 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose: To examine the changes in distraction force following relining of a conventional abdominal aortic stent-graft with a type IIIb endoleak using the Nellix endovascular sealing device compared to a unilateral stent-graft. Methods: Relining is often used to repair type IIIb endoleaks, but the consequences to graft stability are unknown. A mathematical model was constructed based on pressure and volume flow through the stent-grafts, incorporating recognized distraction force equations. Steady flow was presumed at peak systolic pressures to calculate the maximum distraction force, with gravity ignored. Distraction forces for 28- to 36-mm-diameter stent-graft bodies with 16-mm limbs were calculated and compared to forces following relining with single and double Nellix devices or the Renu unilateral device. Results: Distraction forces for the 28-, 32-, and 36-mm stent-grafts prior to relining were 5.99, 10.21, and 14.99 N, respectively. Similar forces were reported after relining with bilateral Nellix devices (5.86, 10.08, and 14.86 N, respectively). However, use of a unilateral Nellix increased the distraction forces to 9.92, 14.14, and 18.92 N, respectively. These were comparable to the increase observed after relining with a Renu unilateral stent-graft (9.87, 14.09, and 18.86 N, respectively). The proportional increase in distraction force for a unilateral relining ranged from 26% to 66%, with the greatest increase noted in the smaller diameter main bodies. Conclusion: Relining a stent-graft with a type IIIb endoleak using bilateral Nellix devices does not increase the distraction force. However, a unilateral Nellix device or the Renu system could theoretically increase the distraction force by up to 66%, potentially risking migration and type Ia endoleak. In clinical practice, these results suggest that a relining with bilateral Nellix may have benefits over the Renu unilateral stent-graft. |
---|---|
ISSN: | 1526-6028 1545-1550 |
DOI: | 10.1177/1526602816631106 |