Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays
Background Use of a hospital pneumatic tube system may be associated with measurement errors. Methods A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma...
Gespeichert in:
Veröffentlicht in: | Annals of clinical biochemistry 2016-03, Vol.53 (2), p.295-297 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Use of a hospital pneumatic tube system may be associated with measurement errors.
Methods
A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma lactate dehydrogenase concentrations were then assayed.
Results
Lactate dehydrogenase concentrations were overestimated (median bias: 18.8%) when evacuated vacuum lithium heparin tubes were sent by pneumatic tube system. This bias was reduced by bubble-wrapping the standard lithium heparin tube or using Monovette lithium heparin tubes in aspiration mode (median bias: +8.7% and −0.3%, respectively).
Conclusions
Cushioning and aspiration-mode sampling may limit pneumatic tube system-associated overestimation of lactate dehydrogenase concentrations. |
---|---|