Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays

Background Use of a hospital pneumatic tube system may be associated with measurement errors. Methods A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Annals of clinical biochemistry 2016-03, Vol.53 (2), p.295-297
Hauptverfasser: Strubi-Vuillaume, Isabelle, Carlier, Valentine, Obeuf, Catherine, Vasseur, Francis, Maury, J-Claude, Maboudou, Patrice, Mangalaboyi, Jacques, Durocher, Alain, Launay, David, Noel, Christian, Brousseau, Thierry
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Use of a hospital pneumatic tube system may be associated with measurement errors. Methods A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma lactate dehydrogenase concentrations were then assayed. Results Lactate dehydrogenase concentrations were overestimated (median bias: 18.8%) when evacuated vacuum lithium heparin tubes were sent by pneumatic tube system. This bias was reduced by bubble-wrapping the standard lithium heparin tube or using Monovette lithium heparin tubes in aspiration mode (median bias: +8.7% and −0.3%, respectively). Conclusions Cushioning and aspiration-mode sampling may limit pneumatic tube system-associated overestimation of lactate dehydrogenase concentrations.
ISSN:0004-5632
1758-1001
DOI:10.1177/0004563215586600