Dose response characteristics of a novel CCD camera-based electronic portal imaging device comparison with OCTAVIUS detector
Dosimetric properties of a CCD camera-based Electronic Portal Imaging Device (EPID) for clinical dosimetric application have been evaluated. Characteristics obtained by EPID also compared with commercial 2D array of ion chambers. Portal images acquired in dosimetry mode then exported raw fluence or...
Gespeichert in:
Veröffentlicht in: | Journal of cancer research and therapeutics 2015-10, Vol.11 (4), p.765-769 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Dosimetric properties of a CCD camera-based Electronic Portal Imaging Device (EPID) for clinical dosimetric application have been evaluated. Characteristics obtained by EPID also compared with commercial 2D array of ion chambers.
Portal images acquired in dosimetry mode then exported raw fluence or uncorrected images were investigated. Integration time of image acquisition mode has adjusted on 1 s per frame.
As saturation of camera of the EPID, dose response does not have linear behavior. The slight nonlinearity of the camera response can be corrected by a logarithmic expression. A fourth order polynomial regression model with coefficient of determination of 0.998 predicts a response to absolute dose values at less than 50 cGy. A field size dependent response of up to 7% (0.99-1.06) relative OCTAVIUS detector measurement was found. The EPID response can be fitted by a cubic regression for field size changes, yielded coefficient of determination of 0.999.
These results indicate that the EPID is well suited for accurate dosimetric purposes, the major limitation currently being due to integration time and dead-time in frame acquisition. |
---|---|