Computed Tomography Estimation of Cochlear Duct Length Can Predict Full Insertion in Cochlear Implantation
OBJECTIVE:To compare the rates of full insertion of electrodes and hearing outcomes obtained with 28-mm and 31-mm cochlear implant electrode arrays. To assess whether cochlear duct length (CDL) estimated by preoperative computed tomography (CT) predicts whether an electrode is fully inserted. STUDY...
Gespeichert in:
Veröffentlicht in: | Otology & neurotology 2016-03, Vol.37 (3), p.223-228 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | OBJECTIVE:To compare the rates of full insertion of electrodes and hearing outcomes obtained with 28-mm and 31-mm cochlear implant electrode arrays. To assess whether cochlear duct length (CDL) estimated by preoperative computed tomography (CT) predicts whether an electrode is fully inserted.
STUDY DESIGN:A cohort study compared electrodes inserted and hearing outcomes after implantation with 28-mm or 31-mm arrays. CDL estimated from preoperative CT was compared in patients in whom full insertion of the 28-mm array was achieved compared with patients in whom at least one basal electrode was outside the cochlea.
SETTING:Tertiary referral cochlear implantation center.
PATIENTS:One hundred forty-eight patients implanted with 28-mm cochlear implant arrays (175 devices) and 74 patients implanted with 31-mm arrays (88 devices).
INTERVENTION:Cochlear implantation with 28-mm or 31-mm array.
MAIN OUTCOME MEASURESActive electrodes at first programming, and at subsequent follow-ups. Bamford–Kowal–Bench sentence and auditory speech sound evaluation test results at 2 to 3 months in adult patients. CDL predicted by preoperative CT.
RESULTS:There was no difference in full insertion between the 28-mm and 31-mm array cohorts (p = 0.22). Early hearing outcomes at 2 to 3 months showed no difference in mean auditory speech sound evaluation (p = 0.19) or Bamford–Kowal–Bench results (p = 0.853) between the 28-mm and 31-mm cohorts. CDL was shorter in the 22 patients with less than full insertion of the array with a mean length of 28.7 mm compared with 29.6 mm in the 42 patients in whom full insertion was achieved (p = 0.046).
CONCLUSION:CT estimation of CDL predicts full insertion after cochlear implantation. Insertion depth does not affect early hearing outcome. |
---|---|