Polarization characterization on the DEIMOS beamline using dichroism measurements
The DEIMOS (Dichroism Experimental Installation for Magneto-Optical Spectroscopy) beamline is a SOLEIL Phase-II beamline and has been opened to users since March 2011. It delivers polarized soft x-rays to perform x-ray absorption spectroscopy, x-ray magnetic circular dichroism and x-ray linear dichr...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2013-01, Vol.425 (21), p.212007-4 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The DEIMOS (Dichroism Experimental Installation for Magneto-Optical Spectroscopy) beamline is a SOLEIL Phase-II beamline and has been opened to users since March 2011. It delivers polarized soft x-rays to perform x-ray absorption spectroscopy, x-ray magnetic circular dichroism and x-ray linear dichroism in the energy range 350-2500 eV. The beamline has been optimized for stability and reproducibility in terms of photon flux and photon energy. The main end-station consists in a cryo-magnet with 2 split coils allowing 7T along the beam or 2T perpendicular to the beam with a variable temperature on the sample from 370K down to 1.5K. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/425/21/212007 |