Highly undersampled peripheral Time-of-Flight magnetic resonance angiography: optimized data acquisition and iterative image reconstruction
Object The aim of this study was to investigate the acceleration of peripheral Time-of-Flight magnetic resonance angiography using Compressed Sensing and parallel magnetic resonance imaging (MRI) while preserving image quality and vascular contrast. Materials and methods An analytical sampling patte...
Gespeichert in:
Veröffentlicht in: | Magma (New York, N.Y.) N.Y.), 2015-10, Vol.28 (5), p.437-446 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Object
The aim of this study was to investigate the acceleration of peripheral Time-of-Flight magnetic resonance angiography using Compressed Sensing and parallel magnetic resonance imaging (MRI) while preserving image quality and vascular contrast.
Materials and methods
An analytical sampling pattern is proposed that combines aspects of parallel MRI and Compressed Sensing. It is used in combination with a dedicated Split Bregman algorithm. This approach is compared with current state-of-the-art patterns and reconstruction algorithms.
Results
The acquisition time was reduced from 30 to 2.5 min in a study using ten volunteer data sets, while showing improved sharpness, better contrast and higher accuracy compared to state-of-the-art techniques.
Conclusion
This study showed the benefits of the proposed dedicated analytical sampling pattern and Split Bregman algorithm for optimizing the Compressed Sensing reconstruction of highly accelerated peripheral Time-of-Flight data. |
---|---|
ISSN: | 0968-5243 1352-8661 |
DOI: | 10.1007/s10334-014-0477-9 |