Performance of a cylindrical diode array for use in a 1.5 T MR-linac

At the UMC Utrecht, a linear accelerator with integrated magnetic resonance imaging (MRI) has been developed, the MR-linac. Patient-specific quality assurance (QA) of treatment plans for MRI-based image guided radiotherapy requires QA equipment compatible with this 1.5 T magnetic field. The purpose...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics in medicine & biology 2016-02, Vol.61 (3), p.N80-N89
Hauptverfasser: Houweling, A C, de Vries, J H W, Wolthaus, J, Woodings, S, Kok, J G M, van Asselen, B, Smit, K, Bel, A, Lagendijk, J J W, Raaymakers, B W
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:At the UMC Utrecht, a linear accelerator with integrated magnetic resonance imaging (MRI) has been developed, the MR-linac. Patient-specific quality assurance (QA) of treatment plans for MRI-based image guided radiotherapy requires QA equipment compatible with this 1.5 T magnetic field. The purpose of this study was to examine the performance characteristics of the ArcCHECK-MR in a transverse 1.5 T magnetic field. To this end, the short-term reproducibility, dose linearity, dose rate dependence, field size dependence, dose per pulse dependence and inter-diode dose response variation of the ArcCHECK-MR diode array were evaluated on a conventional linac and on the MR-linac. The ArcCHECK-MR diode array performed well for all tests on both linacs, no significant differences in performance characteristics were observed. Differences in the maximum dose deviations between both linacs were less than 1.5%. Therefore, we conclude that the ArcCHECK-MR can be used in a transverse 1.5 T magnetic field.
ISSN:0031-9155
1361-6560
DOI:10.1088/0031-9155/61/3/N80