128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose
Objective To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction. Methods Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessme...
Gespeichert in:
Veröffentlicht in: | European radiology 2016-02, Vol.26 (2), p.359-369 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction.
Methods
Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessments were performed. Radiation dose was assessed and compared with the dose of patients who underwent ECG-gated CTAA (n = 126) and the dose of previous CTAA performed with another CT (n = 339).
Results
Image quality (aortic root-ascending portion) was average-to-excellent in more than 94 % of cases, without any non-diagnostic scan. For proximal coronaries, image quality was average-to-excellent in more than 50 %, with only 21.5 % of non-diagnostic cases. Quantitative analysis results were also good. Mean radiation dose for thoracic CTAA was 5.6 mSv versus 20.6 mSv of ECG-gated protocol and 20.6 mSv of 16-slice CTAA scans, with an average dose reduction of 72.8 % (p |
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-015-3848-3 |