128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose

Objective To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction. Methods Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessme...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:European radiology 2016-02, Vol.26 (2), p.359-369
Hauptverfasser: Russo, Vincenzo, Garattoni, Monica, Buia, Francesco, Attinà, Domenico, Lovato, Luigi, Zompatori, Maurizio
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objective To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction. Methods Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessments were performed. Radiation dose was assessed and compared with the dose of patients who underwent ECG-gated CTAA (n = 126) and the dose of previous CTAA performed with another CT (n = 339). Results Image quality (aortic root-ascending portion) was average-to-excellent in more than 94 % of cases, without any non-diagnostic scan. For proximal coronaries, image quality was average-to-excellent in more than 50 %, with only 21.5 % of non-diagnostic cases. Quantitative analysis results were also good. Mean radiation dose for thoracic CTAA was 5.6 mSv versus 20.6 mSv of ECG-gated protocol and 20.6 mSv of 16-slice CTAA scans, with an average dose reduction of 72.8 % (p 
ISSN:0938-7994
1432-1084
DOI:10.1007/s00330-015-3848-3