Using a dose-area product for absolute measurements in small fields: a feasibility study
To extend the dosimetric reference system to field sizes smaller than 2 cm × 2 cm, the LNE-LNHB laboratory is studying an approach based on a new dosimetric quantity named the dose-area product instead of the commonly used absorbed dose at a point. A graphite calorimeter and a plane parallel ion c...
Gespeichert in:
Veröffentlicht in: | Physics in medicine & biology 2016-01, Vol.61 (2), p.650-662 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To extend the dosimetric reference system to field sizes smaller than 2 cm × 2 cm, the LNE-LNHB laboratory is studying an approach based on a new dosimetric quantity named the dose-area product instead of the commonly used absorbed dose at a point. A graphite calorimeter and a plane parallel ion chamber with a sensitive surface of 3 cm diameter were designed and built for measurements in fields of 2, 1 and 0.75 cm diameter. The detector surface being larger than the beam section, most of the issues linked with absolute dose measurements at a point could be avoided. Calibration factors of the plane parallel ionization chamber were established in terms of dose-area product in water for small fields with an uncertainty smaller than 0.9%. |
---|---|
ISSN: | 0031-9155 1361-6560 |
DOI: | 10.1088/0031-9155/61/2/650 |