Real-world examples of sensitivity failures of the 3%/3mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning
In IMRT/VMAT system commissioning (as with any system), quality is improved by striving for tight tolerances of stringent metrics of accuracy. For 5 cases, passing rates for 3%/3mm gamma analysis were generated following the TG119 instructions. Subsequently, more stringent/sensitive criteria combine...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2013-01, Vol.444 (1), p.12086-5 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In IMRT/VMAT system commissioning (as with any system), quality is improved by striving for tight tolerances of stringent metrics of accuracy. For 5 cases, passing rates for 3%/3mm gamma analysis were generated following the TG119 instructions. Subsequently, more stringent/sensitive criteria combined with advanced volumetric dose analysis were applied, and in each case significant systematic errors were clearly identified despite the high 3%/3mm passing rates. In 4 of 5 cases, the error was easily remedied. These real-world examples of observed "false negatives" (insensitivities) point towards the inappropriateness of the 3%/3mm gamma passing rate metric as the basis for acceptance testing/commissioning of the IMRT/VMAT delivery chain. |
---|---|
ISSN: | 1742-6596 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/444/1/012086 |