Lethal Aorto-Right Ventricular Defect After Transcatheter Aortic Valve Implantation in a Patient With Radiation-Induced Porcelain Aorta: Notes of Caution
Abstract A 47-year-old man with severe radiation-induced aortic stenosis was rejected for cardiac surgery because of porcelain aorta. We successfully implanted an Edwards SAPIEN valve (Edwards Lifesciences, Irvine, CA), but the patient was readmitted 3 weeks later for heart failure with a continuous...
Gespeichert in:
Veröffentlicht in: | Canadian journal of cardiology 2016-01, Vol.32 (1), p.135.e9-135.e11 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract A 47-year-old man with severe radiation-induced aortic stenosis was rejected for cardiac surgery because of porcelain aorta. We successfully implanted an Edwards SAPIEN valve (Edwards Lifesciences, Irvine, CA), but the patient was readmitted 3 weeks later for heart failure with a continuous murmur on auscultation. Echocardiography showed a small defect between the aorta and the infundibulum of the right ventricle, which was also confirmed with aortography and computed tomography. Medical therapy was optimized; however, he died unexpectedly a few weeks later. We concluded that irradiated tissues are particularly fragile and require specific attention during transcatheter aortic valve implantation. Furthermore, this case suggests that a more aggressive closure should have been applied. |
---|---|