Evaluation of Dimensional and Flow Properties of ExPress Glaucoma Drainage Devices

PURPOSE:ExPress devices are available as P50 and P200 models, the numbers related to their luminal diameters in μm. We compared their Poiseuille’s Law-based theoretical resistance values with experimental values and correlated these with their luminal dimensions derived from electron microscopy. MET...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of glaucoma 2016-01, Vol.25 (1), p.e39-e45
Hauptverfasser: Samsudin, Amir, Eames, Ian, Brocchini, Steve, Khaw, Peng Tee
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:PURPOSE:ExPress devices are available as P50 and P200 models, the numbers related to their luminal diameters in μm. We compared their Poiseuille’s Law-based theoretical resistance values with experimental values and correlated these with their luminal dimensions derived from electron microscopy. METHODS:Scanning electron microscopy was performed on P50 and P200 devices. Bench-top flow studies were performed to find the resistances of the devices. Devices were also incorporated into a perfused, ex vivo porcine sclera model to test and compare their control of pressure, with and without overlying scleral flaps, and with trabeculectomies. RESULTS:The luminal dimensions of the P200 device were 206.4±3.3 and 204.5±0.9 μm at the subconjunctival space and anterior chamber ends, respectively. Those of the P50 device were 205.0±5.8 and 206.9±3.7 μm, respectively. There were no significant differences between the P200 and P50 devices (all P>0.05). The resistances of the P200 and P50 devices were 0.010±0.001 and 0.054±0.002 mm Hg/μL/min, respectively (P
ISSN:1057-0829
1536-481X
DOI:10.1097/IJG.0000000000000243