Characterization of miniature RAD-HARD silicon diodes as dosimeters for small fields of photon beams used in radiotherapy
The dosimetric response characterization and beam data acquisition performance of a miniature Float Zone (FZ) silicon diode for photon beams was investigated using Novalis TX linear accelerator (Varian Medical Systems®). In all measurements the unbiased diode operated in a short-circuit mode, connec...
Gespeichert in:
Veröffentlicht in: | Radiation measurements 2014-12, Vol.71, p.396-401 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The dosimetric response characterization and beam data acquisition performance of a miniature Float Zone (FZ) silicon diode for photon beams was investigated using Novalis TX linear accelerator (Varian Medical Systems®). In all measurements the unbiased diode operated in a short-circuit mode, connected to the input of a Keithley 6517B electrometer using a water phantom. For photon beams of the 6 and 15 MV the results presented good repeatability (coefficient of variation ≤1.6%), measured through switching on/off the photon beams. Moreover, the diode showed a quite linear response, given by the charge versus absorbed dose, with charge sensitivities higher than 6.9 nC/Gy. The output factor, percentage depth dose profile (PDD) and transversal dose profile (TDP) were also measured in a water phantom. For small field sizes, the output factor values using the FZ diode were compared with measurements obtained with a SFD (Stereotactic Field Diode) commercial diode and the differences were 5.4%, 2.5% and 1.3% for the field sizes of 1 × 1, 2 × 2 and 3 × 3 cm². For larger field sizes (≥4 × 4 cm²), the maximum difference found was 0.7% in comparison with values obtained with a CC13 ionization chamber. Thus, the result demonstrates that the unshielded FZ diode has the potential to be used for measuring of, as it performed acceptably well for both small and large field sizes. The TDP experimental results obtained with the FZ diode for field sizes of: 1 × 1 cm2, 2 × 2 cm2 and 4 × 4 cm2 are in agreement with experimental results acquired with several commercial detectors. Through the TDP study, the comparison of the field penumbra size confirmed the excellent spatial resolution of the miniature diode. However, the PDD study, requires further investigation.
•Dosimetric response characterization of FZ Si diodes for photon beams was studied.•Beam data acquisition performance of FZ diodes for photon beams was investigated.•The FZ diode presented good short term repeatability operating in short-circuit mode.•The results of the penumbra profiles confirmed the good spatial resolution of the FZ.•Output factor study of the FZ matched with different detectors. |
---|---|
ISSN: | 1350-4487 1879-0925 |
DOI: | 10.1016/j.radmeas.2014.08.002 |