Patient delivery quality assurance for linac-based IMRT and helical tomotherapy using solid state detectors

The purpose of this study is to demonstrate how to perform and define appropriate action levels for the intensity-modulated radiation therapy (IMRT) delivery quality assurance (DQA) for both linear accelerator (linac) and helical tomotherapy (HT) plans using a 2D array diode detector. MapCHECK (Sun...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation measurements 2011-12, Vol.46 (12), p.1993-1995
Hauptverfasser: Chong, Ngot Swan, Lee, Jason J.S., Kung, Wei-Hsiang, Chen, Chien-An, Hsieh, Chen-Hsi, Tien, Hui-Ju, Lin, Chi-Ta, Yeh, Hsin-Pei, Wu, Le-Jung, Huang, Hsien-Da, Shueng, Pei-Wei
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study is to demonstrate how to perform and define appropriate action levels for the intensity-modulated radiation therapy (IMRT) delivery quality assurance (DQA) for both linear accelerator (linac) and helical tomotherapy (HT) plans using a 2D array diode detector. MapCHECK (Sun Nuclear, Melbourne, FL, USA) was used to perform the DQA measurements of patient plans at various sites: brain, head and neck, chest, abdomen, pelvis, spine, bone and others. The measured and planned doses were compared using gamma analysis with a total dose discrepancy of 3% or a gradient difference of 3 mm criteria. DQAs for 507 HT and 380 linac-based IMRT plans were performed, and the mean diode percentage passing rates for 3%/3 mm criteria were 93.8% and 95.1% respectively. For HT delivery, 4.6% and 9.5% of the plans had the passing rates below 80% and 85%, while for linac-based IMRT, 7.4% of the plans had the passing rates lower than 90%. Applying confidence limit with multiplier of 1, action levels for passing rates of 87% and 90% are recommended for HT and IMRT plans respectively. ► Quality assurance was performed for 7–9 fields IMRT and helical tomotherapy plans. ► Confidence limit with multiplier of 1 was used to define the results action levels. ► For helical tomotherapy plans, the overall action level is recommended to be 87%. ► For IMRT linac plans, the recommended overall action level is 90%.
ISSN:1350-4487
1879-0925