Paper diagnostic device for quantitative electrochemical detection of ricin at picomolar levels
We report a paper-based assay platform for detection of ricin a chain. The paper platform is assembled by simple origami paper folding. The sensor is based on quantitative, electrochemical detection of silver nanoparticle labels linked to a magnetic microbead support via a ricin immunosandwich. Impo...
Gespeichert in:
Veröffentlicht in: | Lab on a chip 2015-01, Vol.15 (18), p.3707-3715 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a paper-based assay platform for detection of ricin a chain. The paper platform is assembled by simple origami paper folding. The sensor is based on quantitative, electrochemical detection of silver nanoparticle labels linked to a magnetic microbead support via a ricin immunosandwich. Importantly, ricin was detected at concentrations as low as 34 pM. Additionally, the assay is robust, even in the presence of 100-fold excess hoax materials. Finally, the device is easily remediated after use by incineration. The cost of the device, not including reagents, is just $0.30. The total assay time, including formation of the immunosandwich, is 9.5 min. |
---|---|
ISSN: | 1473-0197 1473-0189 |
DOI: | 10.1039/c5lc00731c |