Use of Cotinine Immunoassay Test Strips for Preclassifying Urine Samples from Smokers and Nonsmokers Prior to Analysis by LC-MS-MS
Cotinine biomarker measurements involving both smokers and nonsmokers must accommodate a broad range of concentrations. Thus, we have routinely preclassified unknown samples as being either “high” or “low” by using an enzyme-linked immunoassay for cotinine prior to analysis by tandem mass spectromet...
Gespeichert in:
Veröffentlicht in: | Journal of analytical toxicology 2005-11, Vol.29 (8), p.814-818 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cotinine biomarker measurements involving both smokers and nonsmokers must accommodate a broad range of concentrations. Thus, we have routinely preclassified unknown samples as being either “high” or “low” by using an enzyme-linked immunoassay for cotinine prior to analysis by tandem mass spectrometry (MS). Although this method is effective, it is also time-consuming and complex; a simpler and faster approach would be useful. Consequently, a screening assay for urine cotinine using an immunochromatographic test strip (NicAlert™) followed by a computerized analysis of the data was examined as a possible alternative. The results indicate that this approach can provide useful classification efficiency when using our target cutoff value of approximately 20 ng/mL. In the analysis of 50 urine samples from nonsmokers with varying degrees of exposure to environmental tobacco smoke, the classification sensitivity and specificity were 88% and 92%, respectively, for cotinine measured by the test strips relative to total cotinine concentrations measured by atmospheric-pressure ionization tandem MS. However, the relatively high cost of the strips may be a limiting factor. |
---|---|
ISSN: | 0146-4760 1945-2403 |