Radon measurements during the building of a low-level laboratory
Radon measurements were provided during the different stages of building of a low-level laboratory in Belgrade. The depth of the laboratory is 12 m, equivalent to 30 m of water with an area of 45 m 2. The whole of the laboratory is hermetically lined with 1 mm Al foil and is ventilated with filtered...
Gespeichert in:
Veröffentlicht in: | Radiation measurements 1999-06, Vol.31 (1), p.371-374 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Radon measurements were provided during the different stages of building of a low-level laboratory in Belgrade. The depth of the laboratory is 12 m, equivalent to 30 m of water with an area of 45 m
2. The whole of the laboratory is hermetically lined with 1 mm Al foil and is ventilated with filtered air. Radon concentrations were measured with the CR-39 detector as well as via the gamma-ray spectroscopic measurements. The radon concentrations in the air were achieved to 20 Bqm
−3 and reduction of secondary and tertiary cosmic-ray fluxes is five times when ventilation, filtering and sealing was applied. |
---|---|
ISSN: | 1350-4487 1879-0925 |