Comparison of the Reliability of Acoustic Radiation Force Impulse Imaging and Supersonic Shear Imaging in Measurement of Liver Stiffness

To compare the reliability of acoustic radiation force impulse (ARFI) imaging and supersonic shear imaging (SSI) in measurement of liver stiffness. This study was approved by the institutional review board, and written informed consent was obtained for all patients. Seventy-nine patients (25 healthy...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiology 2015-12, Vol.277 (3), p.881-886
Hauptverfasser: Woo, Hyunsik, Lee, Jae Young, Yoon, Jeong Hee, Kim, Won, Cho, Belong, Choi, Byung Ihn
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To compare the reliability of acoustic radiation force impulse (ARFI) imaging and supersonic shear imaging (SSI) in measurement of liver stiffness. This study was approved by the institutional review board, and written informed consent was obtained for all patients. Seventy-nine patients (25 healthy patients, 26 with Child-Pugh class A, and 28 with Child-Pugh class B or C) were enrolled and analyzed from April 2012 to April 2013. In each patient, three abdominal radiologists performed nine measurements of hepatic shear-wave speed with both ARFI imaging and SSI on the same day. Four weeks later, a second session was performed with the same protocol. Interobserver and intraobserver agreements were calculated by using intraclass correlation coefficients. Technical failures and measurement time were evaluated. There were four technical failures in the SSI group and one in the ARFI group (P = .375). The overall interobserver agreement of ARFI imaging was significantly higher than that of SSI (0.941 vs 0.828, P < .001). The overall intraobserver agreement of ARFI imaging was significantly higher than that of SSI (0.915 vs 0.829, P < .001). The overall shear-wave speed measured with SSI was higher than that measured with ARFI imaging (2.04 m/sec ± 0.88 vs 1.80 m/sec ± 0.81, P < .001). The measurement time of SSI was longer than that of ARFI imaging (310.8 seconds ± 88.5 vs 84.5 seconds ± 15.4, P < .001). ARFI imaging was more reliable than SSI in measurement of liver stiffness. The hepatic shear-wave speed measured with SSI was higher than that measured with ARFI imaging, which means that the shear-wave speeds measured with ARFI imaging and SSI cannot be used interchangeably.
ISSN:0033-8419
1527-1315
DOI:10.1148/radiol.2015141975