User-reported abrasion-related lead failure is more common with durata compared to other implantable cardiac defibrillator leads
Background Following increased rates of inside-out abrasion with the St Jude Medical Riata lead, the Durata implantable cardiac defibrillator (ICD) lead was introduced with modifications intended to increase abrasion resistance. Recent case reports have described insulation failures of the Durata. O...
Gespeichert in:
Veröffentlicht in: | Heart rhythm 2015-12, Vol.12 (12), p.2376-2380 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background Following increased rates of inside-out abrasion with the St Jude Medical Riata lead, the Durata implantable cardiac defibrillator (ICD) lead was introduced with modifications intended to increase abrasion resistance. Recent case reports have described insulation failures of the Durata. Objective To determine if increased rates of abrasion-related failure are present with the Durata lead. Methods The Food and Drug Administration Manufacturer and User Facility Device Experience database was queried for reports of insulation failure of the Durata lead from 2014. Comparison was made to other ICD leads. Incidence was estimated and case characteristics were compared. Results The estimated incidence of abrasion was significantly higher for the Durata lead than for the Boston Scientific Endotak or the Medtronic Quattro leads. The mode of abrasion was most often lead-to-can, as compared to “inside-out” abrasion with the Riata lead. Full-thickness abrasion was associated with failure to defibrillate or inappropriate therapy. Four patients had failure of therapy or death. Conclusions The findings indicate higher rates of insulation failures of the Durata lead, despite design modifications. External abrasion from the pulse generator to the adjacent lead within the device pocket was the most common etiology. Shocks unmasked previously undetected abrasion, resulting in failure to defibrillate. Data are presented indicating a possible time dependency to abrasion risk. This limited query suggests need for ongoing scrutiny of Durata lead performance. Careful inspection of Durata leads at the time of ICD replacement is warranted, as are vigorous attempts to gather information about terminal events in patients with Durata leads. |
---|---|
ISSN: | 1547-5271 1556-3871 |
DOI: | 10.1016/j.hrthm.2015.07.013 |