A novel Hartman Shack-based topography system: repeatability and agreement for corneal power with Scheimpflug+Placido topographer and rotating prism auto-keratorefractor
The purpose of this study is to analyze the repeatability and agreement of corneal power using a new Hartman type topographer in comparison to Scheimpflug+Placido and autorefractor devices. In this cross sectional, observational study performed at the cornea services of a specialty hospital, 100 nor...
Gespeichert in:
Veröffentlicht in: | International ophthalmology 2015-12, Vol.35 (6), p.869-880 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study is to analyze the repeatability and agreement of corneal power using a new Hartman type topographer in comparison to Scheimpflug+Placido and autorefractor devices. In this cross sectional, observational study performed at the cornea services of a specialty hospital, 100 normal eyes (100 consecutive candidates) without any previous ocular surgery or morbidity except refractive error were evaluated. All candidates underwent three measurements each on a Full gradient, Hartman type topographer (FG) (iDesign, AMO), Scheimpflug+Placido topographer (SP) (Sirius, CSO) and rotating prism auto-keratorefractor (AR) (KR1, Nidek). The parameters assessed were flat keratometry (K1), steep keratometry (K2), steep axis (K2 axis), mean
K
,
J
0
and
J
45
. Intra-device repeatability and inter-device agreement were evaluated. On repeatability analysis, the intra-device means were not significantly different (ANOVA,
p
> 0.05). Intraclass correlations (ICC) were >0.98 except for
J
0
and
J
45
. In terms of intra-measurement standard deviation (Sw), the SP and FG groups fared better than AR group (
p
0.05,
R
2
0.05). ICC ranged from 0.92 to 0.99 (
p
|
---|---|
ISSN: | 0165-5701 1573-2630 |
DOI: | 10.1007/s10792-015-0065-7 |