Biodistribution and estimation of radiation-absorbed doses in humans for 13N-ammonia PET
Objective To evaluate the biodistribution of radiation-absorbed doses of 13 N-ammonia in healthy people. Methods Five healthy human subjects underwent whole-body PET and CT scans after injection of 555–740 MBq of 13 N-ammonia. Five serial dynamic emission scans in each healthy volunteer were acquire...
Gespeichert in:
Veröffentlicht in: | Annals of nuclear medicine 2015-11, Vol.29 (9), p.810-815 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
To evaluate the biodistribution of radiation-absorbed doses of
13
N-ammonia in healthy people.
Methods
Five healthy human subjects underwent whole-body PET and CT scans after injection of 555–740 MBq of
13
N-ammonia. Five serial dynamic emission scans in each healthy volunteer were acquired. Regions of interest were drawn on the CT image and transferred to the corresponding transverse PET slice. Estimates of the radiation-absorbed doses were calculated using the medical internal radiation dosimetry method.
Results
The highest concentrations of
13
N-ammonia were found in the heart and liver, followed by pancreas, brain, spleen and stomach. The highest absorbed organ doses were to the heart wall (7.14E−03 ± 3.63E−03 mGy/MBq) and kidneys (6.02E−03 ± 3.53E−03 mGy/MBq). The effective dose (ED) was 6.58E−03 ± 1.23E−03 mSv/MBq.
Conclusions
With these new estimates for
13
N-ammonia dosimetry, the results for Chinese people were not appreciably different from those of the previous study performed with old devices. As one of the most important myocardial perfusion PET tracers, the whole-body
13
N-ammonia PET appears to be safe for humans, yielding a relatively modest radiation burden that would allow multiple PET studies on the same subject per year. |
---|---|