Immunofluorescence sensor for water analysis
We demonstrated a bulk optical fluorescence based immunosensor capable of multianalyte water analysis. Calibration curves obtained for 2,4-dichlorophenoxyacetic acid (2,4-D) and simazine had detection limits of 0.035 and 0.026 μg l −1, respectively. The sensor is reusable due to its regenerability a...
Gespeichert in:
Veröffentlicht in: | Sensors and actuators. B, Chemical Chemical, 1998-08, Vol.51 (1), p.181-187 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We demonstrated a bulk optical fluorescence based immunosensor capable of multianalyte water analysis. Calibration curves obtained for 2,4-dichlorophenoxyacetic acid (2,4-D) and simazine had detection limits of 0.035 and 0.026 μg l
−1, respectively. The sensor is reusable due to its regenerability and cost effective due to the use of components customary in the trade. A concept for the detection of a varying range of analytes on the same transducer is discussed. In addition, the effect on the sensor response of a thin high index film deposited on the transducer surface is studied. |
---|---|
ISSN: | 0925-4005 1873-3077 |
DOI: | 10.1016/S0925-4005(98)00187-7 |