Use of electrocardiogram (ECG) electrodes for Bioelectrical Impedance Analysis (BIA)
BIA is a safe, noninvasive, portable and relatively inexpensive method of estimating body composition that is practical and suitable for individual use and large-scale studies. However, the cost of the electrodes recommended by some BIA manufacturers is too high for developing countries; where very...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2012-01, Vol.407 (1), p.12008-8 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BIA is a safe, noninvasive, portable and relatively inexpensive method of estimating body composition that is practical and suitable for individual use and large-scale studies. However, the cost of the electrodes recommended by some BIA manufacturers is too high for developing countries; where very often the long and complicated process of importation reduces the time they can be used. The purpose of this study was to evaluate the use of two types of ECG electrodes (2290 and 2228 by 3M®) in BIA measurements to decrease the costs of the test. The results showed that the 2228 ECG electrodes can be used in BIA measurements for adult's body composition assessment. These electrodes are available in the domestic market and their costs are 92% lower than the electrodes recommended by manufacturer. The results show a new cost-benefit relation for BIA method and make this a more accessible tool for individual tests, large-scale researches and studies in the community. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/407/1/012008 |