Note: Design and realization of a portable edge illumination X-ray phase contrast imaging system
We discuss a portable edge illumination x-ray phase contrast imaging system based on compact piezoelectric motors, which enables its transportation to different environments, e.g., hosting different x-ray source technologies. The analysis of images of standard samples reveals an angular sensitivity...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2015-09, Vol.86 (9), p.096102-096102 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We discuss a portable edge illumination x-ray phase contrast imaging system based on compact piezoelectric motors, which enables its transportation to different environments, e.g., hosting different x-ray source technologies. The analysis of images of standard samples reveals an angular sensitivity of 270 ± 6 nrad, which compares well with the 260 ± 10 nrad reported for previous systems based on stepper motors, demonstrating that system portability can be achieved without affecting phase sensitivity. The results can also be considered a test of the performance of the piezoelectric motors, and as such could be of interest to researchers planning their use in other imaging systems. |
---|---|