Low cost "laserless" FTIR spectrometer on the farm for real-time nitrous oxide soil emission measurements
A low-cost Fourier transform infrared (FTIR) instrument was developed where the traditional He-Ne reference laser was replaced by a low-cost linear encoder. An RMS sampling error of less than 20 nm was achieved by oversampling both the interferogram and the encoder signal and then resampling the int...
Gespeichert in:
Veröffentlicht in: | Applied optics (2004) 2013-02, Vol.52 (4), p.B93-B101 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A low-cost Fourier transform infrared (FTIR) instrument was developed where the traditional He-Ne reference laser was replaced by a low-cost linear encoder. An RMS sampling error of less than 20 nm was achieved by oversampling both the interferogram and the encoder signal and then resampling the interferogram using a correction table for the encoder. A gas calibration model was developed for the system, which was chosen to have a stroke length of 21 mm and, thereby, a resolution of 0.4 cm(-1) after apodization. The instrument was mounted on a vehicle and employed in an agricultural field test for measuring soil emissions, in particular nitrous oxide (N(2)O). The concentration of N(2)O was measured with a root mean squared error of 6 ppb. The results compared well with lab-based gas chromatography measurements. |
---|---|
ISSN: | 1559-128X 2155-3165 1539-4522 |
DOI: | 10.1364/AO.52.000B93 |