Comparison of the bending performance of solid and cannulated spinal pedicle screws using finite element analyses and biomechanical tests

Highlights • Bending performances of solid and cannulated spinal pedicle screws were analyzed and compared using finite element analyses and biomechanical tests. • Cylindrical spinal pedicle screws with a cannulated design had significantly poorer bending performances. • Conical spinal pedicle screw...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical engineering & physics 2015-09, Vol.37 (9), p.879-884
Hauptverfasser: Shih, Kao-Shang, Hsu, Ching-Chi, Hou, Sheng-Mou, Yu, Shan-Chuen, Liaw, Chen-Kun
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • Bending performances of solid and cannulated spinal pedicle screws were analyzed and compared using finite element analyses and biomechanical tests. • Cylindrical spinal pedicle screws with a cannulated design had significantly poorer bending performances. • Conical spinal pedicle screws maintained their bending performances, whether they were solid or of a cannulated design. • Adding a pin to the cannulated spinal pedicle screws did not improve the bending performances.