Accuracy Evaluation of Four Blood Glucose Monitoring Systems in Unaltered Blood Samples in the Low Glycemic Range and Blood Samples in the Concentration Range Defined by ISO 15197
Systems for self-monitoring of blood glucose (SMBG) are expected to be accurate enough to provide reliable measurement results. Especially in the low glycemic range, adequate therapeutic decisions based on reliable results can alleviate complications associated with hypoglycemia. The accuracy of fou...
Gespeichert in:
Veröffentlicht in: | Diabetes technology & therapeutics 2015-09, Vol.17 (9), p.625-634 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Systems for self-monitoring of blood glucose (SMBG) are expected to be accurate enough to provide reliable measurement results. Especially in the low glycemic range, adequate therapeutic decisions based on reliable results can alleviate complications associated with hypoglycemia.
The accuracy of four SMBG systems (system 1 was the ACCU-CHEK(®) Aviva [Roche Diagnostics GmbH, Mannheim, Germany], system 2 was the Contour(®) XT [Bayer Consumer Care AG, Basel, Switzerland], system 3 was the GlucoCheck XL [aktivmed GmbH, Augsberg, Germany], and system 4 was the GlucoMen(®) LX PLUS [A. Menarini Diagnostics S.r.l., Florence, Italy]) with three test-strip lots each was evaluated by calculating mean absolute relative differences (MARDs). Two datasets were evaluated: (1) 100 samples with blood glucose concentrations |
---|---|
ISSN: | 1520-9156 1557-8593 |
DOI: | 10.1089/dia.2015.0043 |