Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging

Objective This study assessed whether radiolucent carbon fiber reinforced-polyetheretherketone (CFR-PEEK) intramedullary nails decreased hardware artifact on magnetic resonance imaging (MRI) and computed tomography (CT) in vitro and in an oncologic patient population. Materials and methods In vitro...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Skeletal radiology 2015-09, Vol.44 (9), p.1317-1325
Hauptverfasser: Zimel, Melissa N., Hwang, Sinchun, Riedel, Elyn R., Healey, John H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objective This study assessed whether radiolucent carbon fiber reinforced-polyetheretherketone (CFR-PEEK) intramedullary nails decreased hardware artifact on magnetic resonance imaging (MRI) and computed tomography (CT) in vitro and in an oncologic patient population. Materials and methods In vitro and clinical evaluations were done. A qualitative assessment of metal artifact was performed using CFR-PEEK and titanium nail MRI phantoms. Eight patients with a femoral or tibial prophylactic CFR-PEEK nail were retrospectively identified. All patients had postoperative surveillance imaging by MRI, CT, and were followed for a median 20 months (range, 12–28 months). CFR-PEEK images were compared to images from a comparative group of patients with titanium femoral intramedullary nails who had a postoperative MRI or CT. A musculoskeletal-trained radiologist graded visualization of the cortex, corticomedullary junction, and bone–muscle interface, on T1-weighted (T1W), STIR, and contrast-enhanced T1-weighted fat-saturated (T1W FS) sequences of both groups with a five-point scale, performing independent reviews 4 months apart. Statistical analysis used the Wilcoxon rank-sum test and a weighted kappa. Results Substantially less MRI signal loss occurred in the CFR-PEEK phantom than in the titanium phantom simulation, particularly as the angle increased with respect to direction of the static magnetic field. CFR-PEEK nails had less MRI artifact than titanium nails on scored T1W, STIR, and contrast-enhanced T1W FS MRI sequences ( p  ≤ 0.03). The mean weighted kappa was 0.64, showing excellent intraobserver reliability between readings. Conclusions CFR-PEEK intramedullary nail fixation is a superior alternative to minimize implant artifact on MRI or CT imaging for patients requiring long bone fixation.
ISSN:0364-2348
1432-2161
DOI:10.1007/s00256-015-2158-9