Wear comparison between conventional and highly cross-linked polyethylene against a zirconia head: a concise follow-up, at an average 10 years, of a previous report
We previously reported the results of wear comparison at a minimum of 5 years between highly cross-linked polyethylene (HXLPE) and conventional polyethylene (PE) against a zirconia femoral head. We now report the mean wear at 10 years for 52 patients (56 hips) of the original cohort of 61 patients (...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2013-10, Vol.28 (9), p.1654-1658 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We previously reported the results of wear comparison at a minimum of 5 years between highly cross-linked polyethylene (HXLPE) and conventional polyethylene (PE) against a zirconia femoral head. We now report the mean wear at 10 years for 52 patients (56 hips) of the original cohort of 61 patients (65 hips) who had undergone primary total hip arthroplasty at our hospital between November 1999 and August 2000. The mean steady-state linear wear rate of HXLPE was 0.045 mm/y, compared with 0.080 mm/y for conventional PE (P=.0003). The incidence of osteolysis was 25% in the conventional PE group compared with 0% in the HXLPE group. Our study demonstrated that the steady-state wear rate for HXLPE remains significantly lower than that for conventional PE against a zirconia femoral head at a mean of 10 years after implantation. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2012.12.020 |