Key comparison BIPM.RI(I)-K6 of the standards for absorbed dose to water at 5 g cm super( -2) and 7 g cm super( -2) of the NPL, United Kingdom and the BIPM in accelerator photon beams
A comparison of the dosimetry for accelerator photon beams was carried out between the National Physical Laboratory (NPL) and the Bureau International des Poids et Mesures (BIPM) in two periods from September to November 2013. The comparison was based on the determination of absorbed dose to water f...
Gespeichert in:
Veröffentlicht in: | Metrologia 2015-01, Vol.52, p.1-1 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A comparison of the dosimetry for accelerator photon beams was carried out between the National Physical Laboratory (NPL) and the Bureau International des Poids et Mesures (BIPM) in two periods from September to November 2013. The comparison was based on the determination of absorbed dose to water for three radiation qualities at the NPL. The results, reported as ratios of the NPL and the BIPM evaluations (and with the combined standard uncertainties given in parentheses), are 0.9973(62) at 6 MV, 0.9995(66) at 10 MV and 0.9957(81) at 25 MV. This result is the sixth in the on-going BIPM.RI(I)-K6 series of comparisons. |
---|---|
ISSN: | 0026-1394 1681-7575 |